Description
The Infant Incubator YR06213 is a state-of-the-art neonatal care equipment designed to provide a safe and controlled environment for newborns. This incubator features a 7-inch LCD touch screen control panel that allows for easy monitoring and adjustment of temperature conditions. The air and skin temperature are servo-controlled by a computer to ensure precise maintenance suitable for sensitive neonatal skin. With a humidity control function, a removable humidity trough, and various self-check alarms for added safety, the incubator is engineered for optimal neonatal care. Additionally, the device offers a transfusion shelf, mattress tilting adjustment, guardrail for enhanced safety, and phototherapy as an optional feature.
Market Price
The market price for incubators like the Infant Incubator YR06213 typically ranges between $10,000 and $11,369.6 USD. This range reflects the advanced features and technological innovations integrated into these essential neonatal care devices. For an accurate quote, we invite you to use the Kalstein Plus platform, where you can easily generate an instant quote tailored to your specific requirements.
Frequently Asked Questions
What is the power requirement for the Infant Incubator YR06213? The incubator requires an electrical supply of ~220V 50Hz.
Can the temperature settings be adjusted? Yes, the temperature settings can be adjusted over a range, with a >37°C setting available.
Does the incubator have alarms? Yes, the incubator is equipped with various self-check alarms, including air and skin temperature alarms.
Advantages and Disadvantages
The Infant Incubator YR06213 offers numerous advantages, such as precise temperature and humidity control, user-friendly control via a touch screen, and optional features like phototherapy. However, one potential disadvantage could be its higher price range compared to basic models, reflecting its advanced capabilities and features.
Use of the Product in the Field
In practical scenarios, the Infant Incubator YR06213 is utilized in hospital neonatal care units, where it provides a stable and controlled environment for premature or ill infants. The equipment’s precision in maintaining optimal environmental conditions is crucial for the recovery and development of newborns.
Recommendations
To maximize the performance of the Infant Incubator YR06213, regular maintenance and cleaning are advised, especially of components like the humidity trough which is designed for easy removal and cleaning. Ensure all alarms are properly functioning and routinely calibrate the temperature sensors for accurate readings.
Features
- 7-inch LCD touch screen control panel
- Air temperature and skin temperature servo-controlled by computer
- Various and self-check alarms
- Free step mattress tilting adjustment
- Removable humidity trough, easy to clean
- >37°C temperature setting
- With the transfusion shelf and tray
- Added with guardrail
- RS-232 connector
- Has humidity display and control function
- Phototherapy is optional
Technical Specifications
Model | YR06213 |
Electrical requirement | ~220V 50Hz |
Heater Power | 550VA |
Air mode temperature range | 25.0°C-37.0°C |
Skin mode temperature range | 34.0°C-37.0°C |
Air mode temperature alarm | <38.0°C |
Skin mode temperature alarm | < 40.0°C(Air temperature) |
Air mode high temperature range | 37.1°C~39.0°C |
Skin mode high temperature range | 37.1°C~38.0°C |
High temperature mode alarm | < 40.0°C (Air temperature) |
Air mode temperature display range | 20.0°C-45.0°C |
Skin mode temperature display range | 25.0°C~45.0°C |
Air mode alarm | ±3.0°C |
Skin mode alarm | ±1.0°C |
Skin temperature sensor accuracy | ±0.3°C |
Temperature control accuracy | ≤0.5°C |
Mattress temperature uniformity | ≤0.8°C |
Humidity control range | 30-90%RH |
Humidity display range | 0-99.9%RH |