Description
The Biosafety Cabinet Class II A2 YR0090-A is an advanced laboratory device designed to ensure maximum protection for both the user and the surrounding environment from low to moderate risk biological agents. Its sophisticated air curtain isolation design prevents cross-contamination, making it an ideal choice for applications such as cell and tissue culture or handling samples from infected individuals. This cabinet is constructed from high-quality 304 stainless steel, which is corrosion-resistant and easy to disinfect, ensuring longevity and reliability under rigorous laboratory conditions. The built-in features such as the LED LCD control panel, UV lamp protection device, and DOP detection port enhance its usability and safety.
Market Pricing
In the current market, the price range for Class II A2 biosafety cabinets typically spans from approximately $2,800 to $3,600. This competitive range reflects the high-quality engineering and essential safety features integrated into these products, ensuring both value and performance for laboratory professionals.
Frequently Asked Questions
What is the primary function of the Biosafety Cabinet Class II A2?
The primary function is to protect users, the environment, and samples from exposure to potentially hazardous biological agents through physical containment and filtration.
Is the cabinet easy to maintain?
Yes, the stainless steel construction and seamless design make the cabinet easy to clean and maintain, while its advanced filtration system ensures efficient operation.
Advantages and Disadvantages
Advantages:
- Highly effective protection against contamination.
- Durable and easy-to-clean stainless steel construction.
- Advanced safety features and user-friendly controls.
- No need for external ducting or exhaust systems.
- Compact design suitable for various lab sizes.
Disadvantages:
- Initial investment cost may be higher compared to Class I cabinets.
- Regular maintenance and filter changes required for optimal performance.
Product Use in the Field
The Class II A2 cabinet is widely utilized in medical and research laboratories for tasks involving microbiological work with agents that present moderate risks. Examples include processing specimens from infected animals or patients and performing sterile tasks such as preparing cultures. Its effective containment strategies significantly reduce the risk of airborne infections, making it indispensable in clinical and pharmaceutical research settings.
Recommendations
To maximize the longevity and effectiveness of your Biosafety Cabinet Class II A2, regularly schedule maintenance checks and ensure timely replacement of HEPA filters. Follow protocol for safe operation, keeping the sash at the recommended height and avoiding rapid motion near the air curtain to prevent disturbances. Consider utilizing Kalstein Plus for automatic quotations and detailed product information to facilitate seamless procurement.
Features
- Negative pressure vertical laminar flow with 70% internal circulation and 30% exhaust.
- Movable glass door with positioning and height limit alarm.
- Integrated waterproof sockets and sewage interface in the work area.
- Interior made from high-quality 304 stainless steel, smooth and seamless.
- LED LCD panel control and built-in UV lamp protection device.
- DOP detection port and built-in differential pressure gauge.
- Vertical compact operating surface for easy usage.
- HEPA-filtered work surface, interior walls, and exhaust air for maximum safety.
- Designed for use in various laboratory applications, including cell and tissue culture.
Technical Specifications
Model | YR0090-A | |
Cleanliness grade | Class 100@≥0.5μm (U.S. Federal 209E) | |
Number of colonies | ≤0.5pcs/dish·hour (Φ90mm culture plate) | |
Average wind speed | Inside the door | 0.38±0.025m/s |
Middle | 0.26±0.025m/s | |
Inside | 0.27±0.025m/s | |
Front suction air speed | 0.55m±0.025m/s | |
Noise | ≤65dB(A) | |
Vibration half peak | ≤5μm | |
Power supply | AC 220V/50Hz | |
Maximum power consumption | 600W | |
Weight | 140KG | |
Internal Size (mm)W×D×H | 600×500×520 | |
External Size (mm) W×D×H | 760×650×1230 | |
High efficiency filter specification and quantity | 560×440×50×1 380×380×50×1 |
|
Specification and quantity of fluorescent lamp/UV lamp | 8W×1/20W×1 |